Power Planning: Designing Electricity Backwards from the Device List

The Four-Step Method

  1. Load list: voltage, power (sustained/peak), current and plug type per device — get electrical spec sheets from vendors and register each unit.
  2. Circuit allocation: one circuit per high-draw device; circuit capacity = peak × 1.25 safety factor; lighting and office on separate circuits.
  3. Margin: reserve 20-30% free breaker slots — the second device always arrives earlier than planned.
  4. Coordination points: pipe-and-wire inspection before boards close; power-on testing before devices arrive.

Typical Load Profiles

Device class Electrical character
High-power lasers/hair-removal platforms Single- or three-phase, kilowatt-class sustained, high peak current
IPL/RF units Pulsed loads, sensitive to circuit stability
Chillers Sustained duty + pump start currents
Imaging/analysis Low power but EMC-sensitive (EMC)

The Frequent Questions List

FAQ

Who does the planning?

A three-way loop: fit-out electrician draws, device vendors confirm, the clinic reviews — trusting any single party leaves holes.

Adding a device later — rewire?

Depends on margin: free breaker slots and spare conductor capacity mean just adding a circuit; otherwise the panel upgrades — which is exactly why you planned margin.